BlueCombiX


BlueCatBio BlueCombiX Centrifuge and tipless centrifugal washer in one.

The BlueCatBio Blue®CombiX is one of three configurations of BlueCatBio’s Blue®CatX platform. The platform empties plates by centrifugation rather than aspiration: the plate is spun and the liquid leaves the wells without a needle or tip entering them. Blue®FugeX is the centrifuge configuration, Blue®WasherX the washer, and Blue®CombiX — in BlueCatBio’s words — combines centrifugation and evacuation in a single, cost-effective package. Upgrade kits convert one configuration into another.

BlueCatBio positions the CombiX for flow cytometry and 3D cellular assay sample prep on its Blue®Plate, where every spin-down, wash and reagent-addition step happens in the one instrument, and for tipless sample pooling with Blue®Pool. Dispensing from 7 shared and 3 dedicated liquid inputs, and a carousel plate store, are options.

In Revolution, this driver is a scaffold. It registers the instrument so that a system can be configured against it, and nothing more: it opens no connection and sends no command. The rest of this page is precise about where that line falls.

TypePlate centrifuge and centrifugal (tipless) plate washer in one instrument
Plates, washingAny SBS plate up to 44 mm tall, 6- to 1536-well and deep-well
Plates, centrifugationSandwich assays up to 60 mm tall
OptionsDispensing (7 shared and 3 dedicated liquid inputs); carousel plate store
Instrument automation interfaceStandard API, USB and Ethernet, per BlueCatBio
Revolution transportNone — the driver opens no connection
Revolution controlNone. One diagnostic operation that writes to the Revolution log

BlueCatBio publishes the CombiX’s own figures on the Blue®CatX product page and refers its key specifications to the Blue®FugeX and Blue®WasherX, which the CombiX combines.

Wash platesAny SBS format plate up to 44 mm tall (6w – 1536w, deep well)
Centrifuge platesSandwich assays up to 60 mm tall
Centrifuge (Blue®FugeX)Max 1000 g at 3,000 RPM; acceleration and deceleration 1,000 RPM/s
Residual volume (Blue®WasherX)0.01 µL to 5 µL per well, depending on evacuation method
Outer vs. centre row g-force difference, 96-well (Blue®WasherX)6%, against 43% on the original Blue®Washer
Dispenser (Blue®WasherX)CV <±3%; 8 mL dead volume
Dispensing option7 shared and 3 dedicated liquid inputs
Weight90–95 kg, depending on dispenser options
AutomationPlate load and pick-up by a simple robotic gripper; standard API, USB and Ethernet; automated plate loading available via Blue®Bench
SoftwarePC-based graphical interface with preset methods and a method editor; autonomous mode runs methods from internal memory
SafetyCentrifugation shuts off on excessive vibration; sensor-controlled interlocking door
Power100–240 V, 50/60 Hz, 350 W, auto sensing, 10 A max

The product page also lists platform-wide dimensions and a 32 kg weight, but they cannot describe the CombiX, which BlueCatBio lists at 90–95 kg, so they are left out here. Confirm the installed footprint with BlueCatBio before planning a bench.

The figure that decides integration is plate height. Washing takes plates up to 44 mm; the centrifuge takes sandwich assemblies up to 60 mm. Labware taller than that does not go in.

None that reach the instrument. The driver declares a single operation, Debug, which writes the message it is given to the Revolution device log and does nothing else. Initialising the device writes a log line and likewise talks to no hardware.

In practice that means a schedule cannot start a spin, a wash or a dispense, and cannot read the instrument’s state. Any process that runs while a plate is in the CombiX is started by an operator on the instrument’s own software, not by Revolution — so a schedule that routes a plate through this device and expects a washed or spun plate back is relying on something outside Revolution.

The driver declares no fixed set of locations. Any location a mover uses to place a plate on the CombiX is defined in the system configuration, and Revolution treats it as a plain stop on the plate’s route.

Revolution reaches the instrument only as a configured device. There is no transport, no vendor software in the path and no process model. BlueCatBio states the platform offers a standard API over USB and Ethernet; the Revolution driver does not use it.

  • Wash, spin and dispense methods must be set up and started on the instrument’s own software. A schedule step cannot start one.
  • Plates must be within the published heights: 44 mm for washing, 60 mm for centrifuged sandwich assays.
  • Liquid supplies and waste must be set up and running at the instrument. Revolution cannot check them.
  • Timing must be arranged by hand. Because Revolution cannot ask the instrument whether it is busy, a schedule cannot wait on a run finishing.
PropertyPurpose
—The driver declares no configurable device properties. Only the standard device settings — name and locations — apply.

If you need more of this instrument driven from a schedule, get in touch — the driver is extended on demand.

Operations

These are available from a Run Device Operation step.

Debug

Writes the given message to the device log and does nothing else. This driver is a scaffold awaiting implementation, so no command reaches the instrument.

ParameterTypeDefaultDescription
message string BlueCombiX debug

Raise Error

Makes this device report a failure on demand: the step fails with an error naming the device, and the operator is offered the usual error recovery choices. No command reaches the instrument and nothing moves, and the failure is raised in simulation as well as on a real system, so this is the step to use when rehearsing how a schedule handles a device failure.